Noninvasive respiratory monitoring system based on the piezoceramic transducer's pyroelectric effect

2008 
This paper presents a simple alternative method and system for noninvasive respiratory airflow monitoring. The proposed system uses a piezoceramic transducer to measure respiratory airflow. When a piezoceramic transducer is impacted by respiratory airflow, there is a piezoelectric and a pyroelectric response to pressure and thermal airflow fluctuations. In this study, the selected transducer’s response output is dominated by the pyroelectricity factor. Therefore, the piezoelectric effect is not significant and can be ignored in this study. Using the transducer’s pyroelectricity to measure thermal flow variations, a subject’s respiratory rate and respiratory air volumetric flow rate can be monitored. The proposed system was evaluated for accuracy and response time using quiet and postphysical exertion breathing modes. Using the pneumotach system as a benchmark, the proposed system’s respiratory rate measurement accuracy for the two breathing modes is approximately 98.78%. In addition, the proposed system’s...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    16
    Citations
    NaN
    KQI
    []