A 300-MHz Bandwidth Balanced Homodyne Detector for Continuous Variable Quantum Key Distribution
2013
In Gaussian-modulated coherent-state quantum key distribution, the measurement of quadratures of coherent states is performed by using a homodyne detector. However, conventional detectors usually suffer from narrow bands. We present a method to design a high-speed shot-noise-limited balanced homodyne detector. A 300-MHz bandwidth detector is experimentally tested and the level of shot noise is 14 dB higher than the electronic noise. The results show that a detector with this method is potential to design a GHz bandwidth detector for continuous variable quantum key distribution at a low level of ratio of shot noise to electronic noise.