Off-resonance correction of non-Cartesian SWI using internal field map estimation

2021 
Patient-induced inhomogeneities in the magnetic field cause distortions and blurring during acquisitions with long echo times, as in susceptibility-weighted imaging. Most correction methods require collecting an additional ΔB0 field map. To avoid that, we propose a method to approximate this field map using the single echo acquisition only. The main component of the observed phase is linearly related to ΔB0 and TE, and the relative impact of non-ΔB0 terms becomes insignificant with TE>20ms at 3T. The estimated 3D field maps, produced at 0.6 mm isotropic under 3 minutes, provide equivalent corrections to acquired ones.