Technical Note: Dosimetric Characterization of the Dynamic Beam Flattening MLC Sequence on a Ring Shaped, Jawless Linear Accelerator with Double Stacked MLC

2019 
PURPOSE: To characterize the dosimetric features and limitations of the dynamic beam flattening (DBF) on the Halcyon 2.0 linear accelerator (Varian Medical Systems). METHODS: A predefined multi-leaf collimator (MLC) sequence was introduced and used to flatten the 6 MV flattening filter free (FFF) beam on the Halcyon 2.0. Dosimetric characterizations of the flattened beams, including beam flatness, symmetry, percent depth dose (PDD), output factor and MU linearity, were investigated. Flatness and symmetry were obtained from profile measurements with both radiographic films (EDR2) and a two dimensional ion-chamber array (IC Profiler, Sun Nuclear Corporation). MU linearity, output factors, and PDDs were measured in a water tank with a CC13 ion chamber (Scanditronix Wellhofer, Nuremburg, Germany). In addition, the effect of the DBF sequence on 3D plan quality was evaluated by creating DBF plans for a 4-field box rectum and an AP/PA spine plan. Patient specific QA was performed on these plans. RESULTS: At 100 cm SSD and 10 cm depth, a flatness of /=10 cm with DBF. For both larger and smaller field sizes the flatness showed a tendency to increase as the fields got bigger or smaller, respectively. Similar trends in flatness were observed at all depths measured. All measured output factors for square field sizes >/=5 cm were within 1% of the TPS prediction. Linearity was