Determination of Trantinterol Enantiomers in Human Plasma by High-Performance Liquid Chromatography – Tandem Mass Spectrometry Using Vancomycin Chiral Stationary Phase and Solid Phase Extraction and Stereoselective Pharmacokinetic Application

2015 
A sensitive and enantioselective vancomycin chiral stationary phase high-performance liquid chromatography–tandem mass spectrometry method was developed for the determination of trantinterol enantiomers in human plasma. Baseline resolution was achieved using the vancomycin chiral stationary phase known as Chirobiotic V with polar ionic mobile phase consisting of acetonitrile–methanol (60:40, v/v) containing 0.01% ammonia and 0.02% acetic acid at a flow rate of 1.0 mL/min. Waters Oasis HLB C18 solid phase extraction cartridges were used in the sample preparation of trantinterol samples from plasma. The detection was performed on a triple-quadrupole tandem mass spectrometer by multiple reaction monitoring mode via electrospray ionization. The calibration curve was linear in a concentration range from 0.0606 to 30.3 ng/mL in plasma, with the lower limit of quantification of 0.0606 ng/mL. The intra- and interday precision (relative standard deviation) values were within 9.7% and the accuracy (relative error) was from −6.6 to 7.2% at all quality control levels. The method was successfully applied to a study of stereoselective pharmacokinetics in human. Chirality 27:327–331, 2015.© 2015 Wiley Periodicals, Inc.