Multicentre Evaluation of MRI‐based Radiomics Features: A Phantom Study

2020 
INTRODUCTION: This work describes the development of a novel radiomics phantom designed for MRI that can be used in a multi-centre setting. The purpose of this study is to assess the stability and reproducibility of MRI-based radiomics features using this phantom across different MRI scanners. METHODS & MATERIALS: A set of phantoms were 3D printed using MRI visible materials. One set of phantoms were imaged on seven MRI scanners and one was imaged on one MRI scanner. Radiomics analysis of the phantoms, which included first order features, shape and texture features was performed. Intraclass correlation coefficient (ICC) was used to assess the stability of radiomics features across 8 scanners and the reproducibility of two printed models on one scanner. Coefficient of variation (COV) was used to assess the reproducibility of radiomics measurements in the phantom on a single scanner. RESULTS: The phantom models provide sufficient signal-to-noise and contrast in all the tumor models permitting robust automatic segmentation. During a 12-month period of monitoring, the phantom material was stable with T1 and T2 of 150.7 +/- 6.7 ms and 56.1 +/- 3.9 ms, respectively. Of all the radiomics features computed, 34/69 had COV less than 10%. Features from first order statistics were the most robust in stability across the eight scanners with 8/12 (67%) having high stability. 29/50 (58%) texture features had high stability and no shape features had high stability features across the eight scanners. CONCLUSION: A novel MRI radiomics phantom has been developed to assess the reproducibility and stability of MRI-based radiomics features across multiple institutions. The variation in radiomic feature stability demonstrates the need for caution when interpreting these features for clinical studies.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    52
    References
    12
    Citations
    NaN
    KQI
    []