Implementation of a Cost-Effective Unlabeled Probe High-Resolution Melt Assay for Genotyping of Factor V Leiden

2011 
The Factor V Leiden mutation (FVL; c.1601G>A, p.Arg534Gln), the most common aberration underlying activated Protein C resistance, results in disruption of a major anticoagulation pathway and is a leading cause of inherited thrombophilia. A high-throughput assay for FVL mutation detection was developed using a single unlabeled probe on a high-resolution platform, the 96-well Roche 480 LightCycler (LC480) instrument. This method replaced the U.S. Food and Drug Administration–approved Roche Factor V Leiden kit assay on the LightCycler PCR instrument, decreasing total cost by 48%. The analytical sensitivity and specificity of the LC480 high-resolution assay approached 100% for the FVL mutation. Factor V mutations in proximity to the FVL locus may influence probe binding efficiency and melt characteristics. One out of three very rare variants tested in a separate study, 1600delC, was not distinguishable from FVL using the described high-resolution assay. However, a c.1598G>A variant, which changes the amino ac...