The effect of different CT scanners, scan parameters and scanning setup on Hounsfield units and calibrated bone density: a phantom study

2018 
This study investigated the effect of different CT scanners, reconstruction protocols, scan positions, and air gaps on HU and/or calibrated calcium hydroxyapatite concentrations (CaHA). Phantoms containing bone-like materials were scanned on four different CT scanners. The effect of slice thickness, field of view (FOV) and reconstruction kernel on HU was investigated. Using clinical scan and reconstruction protocols and a calibration phantom, HU and CaHA were determined for different positions, and air gaps between phantom and calibration phantom. Changing reconstruction kernel considerably affected the HU (range -31 HU to 64 HU), whereas slice thickness and FOV did not. Inter-scanner differences in HU were