Continuous-Time Acquisition of Biosignals Using a Charge-Based ADC Topology

2018 
This paper investigates continuous-time (CT) signal acquisition as an activity-dependent and nonuniform sampling alternative to conventional fixed-rate digitisation. We demonstrate the applicability to biosignal representation by quantifying the achievable bandwidth saving by nonuniform quantisation to commonly recorded biological signal fragments allowing a compression ratio of $\approx$ 5 and 26 when applied to electrocardiogram and extracellular action potential signals, respectively. We describe several desirable properties of CT sampling, including bandwidth reduction, elimination/reduction of quantisation error, and describe its impact on aliasing. This is followed by demonstration of a resource-efficient hardware implementation. We propose a novel circuit topology for a charge-based CT analogue-to-digital converter that has been optimized for the acquisition of neural signals. This has been implemented in a commercially available 0.35  $\mu \text{m}$ CMOS technology occupying a compact footprint of 0.12 mm 2 . Silicon verified measurements demonstrate an 8-bit resolution and a 4 kHz bandwidth with static power consumption of 3.75  $\mu$ W from a 1.5 V supply. The dynamic power dissipation is completely activity-dependent, requiring 1.39 pJ energy per conversion.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    38
    References
    3
    Citations
    NaN
    KQI
    []