Visualization of radiotracers for SPECT imaging using a Timepix detector with a coded aperture

2020 
The work shows the ability to visualize radiotracers used in SPECT with a system based on a coded aperture mask and a hybrid pixel Timepix detector with the CdTe sensor. Characterization of the system using X-rays and radioactive sources confirms that the spatial resolution of less than 1 mm with a field of view 3 cm x 3 cm can be achieved. The results of a simulation study to determine the expected spatial resolution of the system in the focal plane for the various radionuclides is presented. The possibility of using this system with a thin (1.5 mm) coded aperture mask for reconstructing images of gamma emitters with the energy up to 180 keV is demonstrated.