An investigation of the accuracy and reliability of body composition assessed with a handheld electrical impedance myography device

2018 
AbstractThe purpose of this study was to compare the body fat per cent (BF%) assessed with a unique handheld electrical impedance myography (EIM) device, along with other popular methods, to dual-energy X-ray absorptiometry (DXA). Participants included 33 males (aged 24.3 ± 4.6 years) and 38 females (aged 25.3 ± 8.9 years) who completed 2 visits separated by 24–72 h. The assessments included DXA, bioelectrical impedance analysis (BIA), skinfold measures (SKF), and three separate EIM measurements. No significant differences in BF% (P > 0.05) were found between all EIM assessments when compared against DXA for both males and females for each visit. All methods showed no significant differences in BF% (P > 0.05) between days within themselves. Across both days, the standard error of the estimate (SEE) for the EIM measurements ranged from 2.66% to 3.15%, the SEE for BIA was 2.80 and 2.85, and for SKF was 2.90 and 2.82. The 95% limits of agreement ranged from ±5.34% to ±6.38% for EIM measurements and were high...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    26
    References
    6
    Citations
    NaN
    KQI
    []