Quantitative determination of sodium stearyl fumarate in a tablet formulation from continuous manufacturing using a high-pH reversed-phase HPLC method

2017 
ABSTRACTThis study demonstrates the first use of a reversed-phase (RP) high-performance liquid chromatography method with a high-pH buffer for the analysis of sodium stearyl fumarate (SSF) from a tablet formulation. After examining the retention time and peak shape using various buffer concentrations, buffer pH, and RP stationary phases, an optimized method was established using the XBridge® BEH C18 at high pH. This column was further evaluated for method specificity, accuracy, precision, linearity, stability, and sensitivity. Finally, the method was successfully used as a convenient and robust analytical procedure to accurately quantitate SSF in stratified tablets from a continuous manufacturing process to confirm the excipient uniformity throughout the process.