A new system for the simultaneous measurement of δ13C and δ15N by IRMS and radiocarbon by AMS on gaseous samples: Design features and performances of the gas handling interface

2015 
Abstract We present the general design features and preliminary performances of a new system for the simultaneous AMS- 14 C and IRMS δ 13 C and δ 15 N measurements on samples with masses in the μg range. The system consists of an elemental analyzer (EA), a gas splitting unit (GSU), a IRMS system, a gas handling interface (GHI) and a sputtering ion source capable of accepting gaseous samples. A detailed description of the system and of the control software supporting unattended operation are presented together with the first performance tests carried out by analyzing samples with masses ranging from 8 μgC to 2.4 mgC. The performances of the system were tested in term of stability of the ion beam extracted from the ion source, precision and accuracy of the results by comparing the measured isotopic ratios with those expected for reference materials.