High Kinetic Energy Ion Mobility Spectrometer: Quantitative Analysis of Gas Mixtures with Ion Mobility Spectrometry

2014 
We present a high kinetic energy ion mobility spectrometer (HiKE-IMS) for quantitative gas analysis. Drift tube and reaction tube can be operated at reduced fields up to 110 Td. At such conditions the distribution of reactant ion water clusters is shifted toward smaller clusters. Due to the resulting presence of bare reactant ions (e.g., H3O+) and the kinetic control of the ionization process with decreasing reaction time, unlike conventional IMS, a quantitative detection with ppbv detection limits of low proton affine analytes even in humid gas mixtures containing high proton affine compounds is possible using a direct sample gas inlet. A significantly improved dynamic range compared to conventional IMS is achieved. An incremental change in reduced fields enables the observation of parameters like field dependent ion mobilites or analyte fragmentation. Also, the characteristic of the analyte signal with respect to the reduced reaction field gives insight into the ionization process of the analyte. Thus, ...