Correlation and uncertainties evaluation in backscattering of entrance surface air kerma measurements

2018 
The air kerma measurement is important to verify the applied doses in radiodiagnostic. The literature determines some methods to measure the entrance surface air kerma or entrance surface dose but some of these methods may increase the measurement with the backscattering. Was done setups os measurements to do correlations between them. The expanded uncertainty exceeded 5% for measurements with backscattering, reaching 8.36%, while in situations where the backscattering was avoided, the uncertainty was 3.43%.