EVALUATION OF ACETABULAR CUP POSITION AFTER THE COMPUTER-ASSISTED TOTAL HIP ARTHROPLASTY

2008 
Since the autumn of 2003, a computer-assisted system (VectorVision® Hip, version 2.1, Brain LAB, Germany) has been used to perform total hip arthroplasty (THA) operations in our hospital. In the present study, the postoperative acetabular cup position was evaluated using the records of the system and the data measured from postoperative radiographs.To date, 18 patients have been treated using this-system. We studied the cup inclination and anteversion records in this system recorded in the THA procedures. We also measured the cup inclination and anteversion using postoperative radiographs, according to the method described by Pradhan. The inclination and ante-version were the ‘operative’angles for this system and were the ‘radio graphical’ ones for measuring from the radiographs according to the definition described by Murray.The initial planning of the acetabular cup position was 45° ‘operative’ inclination and 20° ‘operative’ ante-version. From the system records, the average ‘operative’ inclination was...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []