Smoke and mirrors: Ultra-rapid-scan FT-IR spectrometry

1998 
Fourier transform-infrared spectrometers have dominated the marketplace and the experimental literature of vibrational spectroscopy for almost three decades. These versatile instruments have been applied to a wide variety of measurements in both industrial and research settings. There has been, however, an ongoing need for enhanced time resolution. Limitations of time resolution in FT-IR measurements arise from the modulation frequencies intrinsic to the spectral multiplexing. Events which are slower than the minimum scan time, about 40 milliseconds at 4-cm−1 resolution, can be readily monitored with conventional instrumentation. For shorter transients, various step-scan, stroboscopic and asynchronous methods have been demonstrated to provide excellent time resolution, down to nanoseconds, but these approaches are limited to events which can be repeated many times with minimal variations. Some of these methods are also susceptible to low-frequency noise sources. The intrinsic scan time of conventional FT-...