Computational Simulations for Infrared Laser Sealing and Cutting of Blood Vessels

2021 
Blood vessel burst pressures were simulated and predicted for sealing and cutting of vessels in a two-step process, using low (<25 W), medium (∼100 W), and high (200 W) power lasers at a wavelength of 1470 nm. Monte Carlo optical transport, heat transfer, Arrhenius integral tissue damage simulations, and vessel pressure equations were utilized. The purpose of these studies was to first validate the numerical model by comparison with experimental results (for low and medium power) and then to use the model to simulate parameters that could not be experimentally tested (for high power). The goal was to reduce the large range of parameters (power, irradiation time, and linear beam dimensions) to be tested in future experiments, for achieving short vessel sealing/cutting times, minimal bifurcated seal zones (BSZ), and high vessel burst pressures. Blood vessels were compressed to 400 μm thickness. A wide range of linear beam profiles (1-5 mm widths and 8-9.5 mm lengths), incident powers (20-200 W) and clinically relevant irradiation times (0.5-5.0 s) were simulated and peak seal and cut temperatures as well as thermal seal zones, ablation zones, and BSZ computed. A simplistic mathematical expression was used to estimate vessel burst pressures based on seal width. Optimal low-power parameters were: 24W/5s/8 × 2 mm (sealing) and 24W/5s/8x1mm (cutting), yielding a BSZ of 0.4 mm, corresponding to experimental burst pressures of ∼450 mmHg. Optimal medium-power parameters were: 90W/1s/9.5x3mm (sealing) and 90W/1s/9.5x1mm (cutting), yielding a BSZ of 0.9 mm for burst pressures of ∼1300 mmHg. Simulated only optimal high-power parameters were: 200W/0.5s/9x3 mm (sealing) and 200W/0.5s/9x1mm (cutting), yielding a BSZ of 0.9 mm and extrapolated to predict a seal strength of ∼1300 mmHg. All lasers produced seal zones between 0.4-1.5 mm, corresponding to high vessel burst pressures of 300-1300 mmHg (well above normal systolic blood pressure of 120 mmHg). Higher laser powers enable shorter sealing/cutting times and higher vessel strengths.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    40
    References
    0
    Citations
    NaN
    KQI
    []