We evaluated the EMIT Cyclosporine Assay, designed for specific quantification of cyclosporine (CsA) in whole blood. The assay was run on the Roche Cobas Mira or Cobas Mira S analyzer. Analytical recovery from both normal donor whole blood and transplant patients' samples was within 17% of the standards. Measurement of diluted out-of-range samples also yielded excellent recovery (101-105%). Within-run, between-run, and total CVs were 4 h, and assay calibration was stable for at least 2 weeks. Patients' samples analyzed by the EMIT assay showed strong correlation with both HPLC and 125I-RIA (r > 0.97). We conclude that the EMIT Cyclosporine Assay provides a convenient, accurate, and precise method for specific quantification of CsA in whole blood.