Comparison of Corneal Tomography: Repeatability, Precision, Misalignment, Mean Elevation, and Mean Pachymetry.
2018
ABSTRACTPurpose: To compare corneal tomography and its statistical uncertainty for measurements obtained by three clinically used corneal tomographers: A Scheimpflug camera (Pentacam HR), a swept source optical coherence tomography system (CASIA SS-1000), and Placido ring imaging (TMS-5).Material and Methods: Repeated measurements with all three devices on 34 normal eyes were used to estimate the repeatability, precision, and mean values of corneal elevation and pachymetry within 8 mm diameter. The repeatability (standard deviation) was calculated for each data point of the corneal elevation data-maps of anterior and posterior cornea as well as for the pachymetry data-maps. Uncertainty on the position of the eye at each measurement might contribute to the differences between elevation data-maps. To take this into account, we defined the precision as the standard deviation for the elevation data-maps of anterior and posterior cornea after correction of misalignment-effects (rotation, translation). The mean...