High-performance liquid chromatography electrospray ionization tandem mass spectrometry for the detection and quantitation of benzo[a]pyrene-DNA adducts

2005 
A method, using HPLC combined with electrospray tandem mass spectrometry (ES-MS/ MS), was developed and validated to detect and quantify the major DNA adduct resulting from exposure to the ultimate tumorigenic benzo[a]pyrene (BP) metabolite, trans-7,8-dihydroxy-anti-9,10-epoxy-7,8,9,10-tetrahydrobenzo[a]pyrene (BPDE). Calf thymus DNA was reacted with BPDE, digested enzymatically to nucleosides, and the major DNA adduct, 10-(deoxyguanosin-N 2 -yl)-7,8,9-trihydroxy-7,8,9,10-tetrahydrobenzo[a]pyrene (dG-BPDE), was purified by HPLC. Similar procedures were applied to prepare dG-BPDE-d 8 from [1,2,3,4,5,6,11,12- 2 H 8 ]BPDE for use as an internal standard. The HPLC-ES-MS/MS method was validated using a mixture of hydrolyzed salmon testis DNA (82 ,ug) and 10 pg dG-BPDE (analogous to 6.9 adducts/10 8 nucleotides). The results indicated an inter- and intraday accuracy of 99-100% and precision of 1.6-1.7% (relative standard deviation). When applied to a calf thymus DNA sample modified in vitro with [1,3- 3 H]BPDE, the method gave a value very similar to those obtained by radiolabeling, 3 2 P-postlabeling, and immunoassay. HPLC-ES-MS/MS analysis of hepatic DNA from mice treated intraperitoneally with 0.5 and 1.0 mg of [7,8- 3 H] BP gave values comparable to those determined by 3 2 P-postlabeling and immunoassay. Lung DNA from mice fed a 0.3% coal tar diet (containing approximately 2 mg BP/g coal tar) for one month had 0.6 ′ 0.04 dG-BPDE adducts/108 nucleotides. This value is much lower than the 102 ′ 14 total DNA adducts/ 10 8 nucleotides determined by 3 2 P-postiabeling, which suggests that dG-BPDE makes only a minor contribution to the DNA adducts formed in lung tissue of mice administered coal tar. The HPLC-ES-MS/MS method was used to assess human lung DNA samples for the presence of dG-BPDE. Based upon a limit of detection of 0.3 dG-BPDE adducts/108 nucleotides, when using 100 μg of DNA, dG-BPDE was detected in only 1 out of 26 samples. These observations indicate that HPLC-ES-MS/MS is suitable to assess the contribution of BP to DNA damage caused by exposures to polycyclic aromatic hydrocarbon (PAH) mixtures. The results further suggest that dG-BPDE may contribute only a small fraction of the total DNA adducts detected by other DNA adduct methodologies in individuals exposed to PAHs.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    10
    References
    84
    Citations
    NaN
    KQI
    []