Development of Smart-Ring-Based Chest Compression Depth Feedback Device for High Quality Chest Compressions: A Proof-of-Concept Study

2021 
Recently, a smart-device-based chest compression depth (CCD) feedback system that helps ensure that chest compressions have adequate depth during cardiopulmonary resuscitation (CPR) was developed. However, no CCD feedback device has been developed for infants, and many feedback systems are inconvenient to use. In this paper, we report the development of a smart-ring-based CCD feedback device for CPR based on an inertial measurement unit, and propose a high-quality chest compression depth estimation algorithm that considers the orientation of the device. The performance of the proposed feedback system was evaluated by comparing it with a linear variable differential transformer in three CPR situations. The experimental results showed compression depth errors of 2.0 ± 1.1, 2.2 ± 0.9, and 1.4 ± 1.1 mm in the three situations. In addition, we conducted a pilot test with an adult/infant mannequin. The results of the experiments show that the proposed smart-ring-based CCD feedback system is applicable to various chest compression methods based on real CPR situations.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    31
    References
    2
    Citations
    NaN
    KQI
    []