Advanced time-stamped total data acquisition control front-end for MeV ion beam microscopy and proton beam writing

2013 
Many ion-matter interactions exhibit [email protected] time dependences such as, fluorophore emission quenching and ion beam induced charge (IBIC). Conventional event-mode MeV ion microbeam data acquisition systems discard the time information. Here we describe a fast time-stamping data acquisition front-end based on the concurrent processing capabilities of a Field Programmable Gate Array (FPGA). The system is intended for MeV ion microscopy and MeV ion beam lithography. The speed of the system (>240,000 events s^-^1 for four analogue to digital converters (ADC)) is limited by the ADC throughput and data handling speed of the host computer.