The μSR setup on the muon beam of the synchrocyclotron at the Konstantinov Institute of Nuclear Physics

2007 
The μSR setup for investigating the distribution of magnetic fields in solids using the muon spin rotation (μSR) method is described. The setup is characterized by a high degree of homogeneity of the magnetic field at the site of the sample under investigation, compensation of scattered magnetic fields to a level of −10−2 G, and a time resolution of 2.5 ns (the full width at half-maximum). The setup is suitable for μSR measurements on samples in the temperature range of 5–300 K with a precision of ±0.1 K.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    1
    References
    6
    Citations
    NaN
    KQI
    []