Automatic Quality Assessment of a Forced Expiratory Manoeuvre Acquired with the Tablet Microphone
2020
Evaluation of lung function is central to the management of chronic obstructive respiratory diseases. It is typically evaluated with a spirometer by a specialized health professional, who ensures the correct execution of a forced expiratory manoeuvre (FEM). Audio recording of a FEM using a smart device embedded microphone can be used to self-monitor lung function between clinical visits. The challenge of microphone spirometry is to ensure the validity and reliability of the FEM, in the absence of a health professional. In particular, the absence of a mouthpiece may allow excessive mouth closure, leading to an incorrect manoeuvre. In this work, a strategy to automatically assess the correct execution of the FEM is proposed and validated. Using 498 FEM recordings, both specificity and sensitivity attained were above 90%. This method provides immediate feedback to the user, by grading the manoeuvre in a visual scale, promoting the repetition of the FEM when needed.