Interfacing a Polymer-Based Micromachined Device to a Nanoelectrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometer
2001
Here we report the design, fabrication, and operation of a polymer-based microchip device interfaced to a nanoelectrospray ionization source and a Fourier transform ion cyclotron resonance mass spectrometer. The poly(methyl methacrylate) micromachined device was fabricated using X-ray lithography to produce a network of channels with high aspect ratios. Fabrication of high aspect ratio channels allows for zero dead volume interfaces between the microchip platform and the nanoelectrospray capillary interface. The performance of this device was evaluated with standard peptide and protein samples. High-quality mass spectral data from peptide and proteins (and mixtures thereof) were obtained without any interfering chemical noise from the polymer or the developers and plasticizers used in the fabrication process. Sample cross-contamination is not a problem using this polymer-based microchip device as demonstrated by the sequential analysis of several proteins. The nanoelectrospray source was operated at flow ...