In Situ Detection of Species Relevant to the Carbon Cycle in Seawater with Submersible Potentiometric Probes

2017 
We report on the development of a submersible probe for the simultaneous potentiometric detection of carbonate, calcium, and pH in seawater. All-solid-state electrodes incorporating nanomaterials provide an adequate response time (<10 s), stability (drifts of <0.9 mV h–1), reproducibility (calibration parameter deviation of <0.7%), and accuracy (deviation of <8% compared to reference techniques) for real-time monitoring of seawater using a flow system. The functioning of the deployable prototype was checked in an outdoor mesocosm and via long-term monitoring in Genoa Harbor. The electrodes worked properly for 3 weeks, and the system demonstrated the capability to autonomously operate with routines for repetitive measurements, data storage, and management. In situ profiles observed in Genoa Harbor and Arcachon Bay were validated using on site and ex situ techniques. The validation of in situ-detected carbonate is a challenge because both re-equilibration of the sample with atmospheric CO2 and the use of ap...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    37
    References
    39
    Citations
    NaN