Final report on the supplementary comparison EURAMET.EM-S26: inductance measurements of 100?mH at 1?kHz (EURAMET project 816)

2012 
Within the framework of EURAMET, a comparison has been organized of measurements of self-inductance at 100?mH and 1?kHz. The comparison is identified as EUROMET.EM-S26. Two 100?mH inductance standards, encased in a thermostatic controlled enclosure, were circulated among the participants. The results from 10 out of 16 participants are in agreement with the reference values. A majority of the results agrees within ?50??H/H. The participants reported four different methods to realize the traceability of the unit of inductance. The results from these different methods are in good agreement within the reported uncertainties. The results of this comparison have been linked to the results of the comparison EUROMET.EM-S20. The organization of this comparison has been a successful exercise of sharing the workload among four different institutes and can be an example to future comparisons. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by EUROMET, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).