Determination of cisapride and norcisapride in human plasma using high-performance liquid chromatography with ultraviolet detection

1998 
Abstract A simple, rapid and reproducible high-performance liquid chromatographic assay for cisapride and norcisapride in human plasma is described. Samples of plasma (150 μl) were extracted using a C 18 solid-phase cartridge. Regenerated tubes were eluted with 1.0 ml of methanol, dried, redissolved in 150 μl of methanol and injected. Chromatography was performed at room temperature by pumping acetonitrile–methanol–0.015 M phosphate buffer pH 2.2–2.3 (680:194:126, v/v/v) at 0.8 ml/min through a C 18 reversed-phase column. Cisapride, norcisapride and internal standard were detected by absorbance at 276 nm and were eluted at 4.3, 5.3 and 8.1 min, respectively. Calibration plots in plasma were linear ( r >0.998) from 10 to 150 ng/ml. Intraday precisions for cisapride and norcisapride were 3.3% and 5.4%, respectively. Interday precisions for cisapride and norcisapride were 9.6% and 9.0%, respectively. Drugs used which might be coadministered were tested for interference.