Cloud Point Extraction and Spectrophotometry Determination of Lead in Environment Water Using 5-Br-PADAP

2013 
A simple, sensitive, green and low cost detection method based on the cloud point extraction (CPE) separation and spectrophotometry was proposed for the determination of lead. In pH=9.0 H3BO3 buffer solution, Pb(II) reacts with 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (5-Br-PADAP) in the presence of Triton X-100 yielding a hydrophobic complex, which then is extracted into micro-volume surfactant-rich phase. The calibration graph was linear in the range of 20-400 µg/L (at 560 nm). Under the optimized conditions, the detection limits of 10.94 µg/L and the relative standard deviations(RSD) of 2.0% (n=5) for Lead(II) were found, respectively. The sensitivity and absorbance of this method are at least five times higher when compared with that of usual 5-Br-PADAP spectrophotometry without CPE, and the proposed method has been applied to the determination of Lead in environment water samples with satisfactory results.