Measurement of (124)I.
2016
Abstract The paper describes the measurements performed at IFIN-HH regarding the creation of a Romanian 124 I standard, consisting of: absolute standardization of the solution by the application of the 4πβ(PC)-γ coincidence method; Calibration of the CENTRONIC IG12/20A ionization chamber with a standardised solution and comparison with a calculated efficiency; γ-ray spectrometry activity measurement and determination of the impurity levels; Comparison of the results of the three methods.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
12
References
5
Citations
NaN