Limited diagnostic possibilities for bloodstream infections with broad‐range methods: A promising PCR/electrospray ionization‐mass spectrometry platform is no longer available

2020 
Fast and accurate detection of causative agents of bloodstream infections remains a challenge of today's microbiology. We compared the performance of cutting-edge technology based on polymerase chain reaction coupled with electrospray ionization-mass spectrometry (PCR/ESI-MS) with that of conventional broad-range 16S rRNA PCR and blood culture to address the current diagnostic possibilities for bloodstream infections. Of 160 blood samples tested, PCR/ESI-MS revealed clinically meaningful microbiological agents in 47 samples that were missed by conventional diagnostic approaches (29.4% of all analyzed samples). Notably, PCR/ESI-MS shortened the time to positivity of the blood culture-positive samples by an average of 34 hr. PCR/ESI-MS technology substantially improved current diagnostic tools and represented an opportunity to make bloodstream infections diagnostics sensitive, accurate, and timely with a broad spectrum of microorganisms covered.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    10
    References
    2
    Citations
    NaN
    KQI
    []