Validation of the single-platform ISHAGE method for CD34+ hematopoietic stem and progenitor cell enumeration in an international multicenter study

2003 
Background Flow cytometric enumeration of CD34 + hematopoietic stem and progenitor cells (HPC) is the reference point for undertaking apheresis and evaluation of adequacy for PBSC engraftment. An external quality assurance (EQA) scheme for CD34 + HPC enumeration has been operational in Belgium, Netherlands and Luxemburg (Benelux) since 1995. Within this group, a multicenter survey was held to validate the state-of-the-art methodology, i.e., multiparametric defini tion of HPC based on light scatter, expression ofCD34 and CD45, and counting beads (i.e., ‘single platform ISHAGE’ method). Methods ‘Real-time’ EQA was used to monitor the application of the single-platform ISHAGE method by 36 participants. Three send-outs of stabilized blood with CD34 + cell counts 35–60 cells/μL were distributed to 36 participants, who were required to assay the samples on three occasions using the standard assay and their local techniques. These results were compared with those obtained by 111–116 UK NEQAS participants testing the same specimens. Results Using the single platform ISHAGE method, between-laboratory coefficients of variation (CVs) as low as 10% were achieved. Intra-laboratory CVs were Conclusion The single-platform ISHAGE method for CD34 + cell enumeration has been validated by an international group of 36 laboratories. The observed variation between laboratories allows a meaningful compar ison of CD34 + cell enumeration results.