Physical performances and kinetics of evaporation of the CIP 10-M personal sampler's rotating cup containing aqueous or viscous collection fluid

2016 
ABSTRACTThe CIP 10-M personal sampler measures worker exposure to airborne particles by collecting particles in a rotating metal cup containing a few milliliters of a collection fluid. This device is mainly used to sample microorganisms or microbial components to measure bioaerosol concentrations in various occupational environments. Aqueous liquids are generally used, but their rapid evaporation limits the duration of sampling; alternative collection fluids could alleviate this problem. Indeed, the particle-collection efficiency of the rotating cup has not been extensively studied, and the only data available relate to a discontinued model. This study aimed to measure the collection efficiency of the current rotating cup model containing an aqueous (water) or viscous (ViaTrap mineral oil) collection fluid. The kinetics of evaporation confirmed that ViaTrap does not evaporate, making 8-h sampling campaigns in constant volumes feasible. Particles with a wide range of aerodynamic diameters (between around 0...