REDUCED CHONDROCYTE VIABILITY IS ASSOCIATED WITH MARKER PEN INK

2012 
Introduction The aim of this study was to investigate whether methylene blue dye, commonly used in sterile surgical marker pens, would have an effect on human chondrocyte viability, when cultured on a collagen membrane in-vitro. Methods Bilayered collagen membranes were seeded in duplicate with 12 million human chondrocytes per ml and cultured for 24 hours under standard conditions. Group A consisted of a membrane marked with methylene blue ink on its smooth side, group B marked on its porous side, and group C acting as an unmarked control. At the end of the culture period the membranes were qualitatively analysed for cell survival by live/dead fluorescent staining under confocal microscopy. Results The control membranes of group C showed normal ‘live’ staining of cells. A marked reduction in chondrocyte viability was seen in groups A and B, with a significant reduction in viability seen opposite to the ink mark on the smooth side in A, and practically no cell viability in the same position on the porous side in group B. Conclusions Methylene blue dye in surgical marker pen ink has been shown to be cytotoxic to chondrocytes cultured on bilayered collagen membranes. Sterile surgical marker pens are commonly used in cartilage repair surgery to aid in the placement of periosteal patches or collagen membranes in autologous chondrocyte transplantation. Surgeons need to be aware of the potential harmful effects of methylene blue dye when using surgical marker pens in cell based regenerative therapies.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []