A novel sequence to improve auditory functional MRI with variable silent delays.

2020 
PURPOSE Auditory functional MRI (fMRI) often uses silent inter-volume delays for stimulus presentation. However, maintaining the steady-state of the magnetization usually requires constant delays. Here, a novel acquisition scheme dubbed "pre-Saturated EPI using Multiple delays in Steady-state" (SEPIMS) is proposed, using spin saturation at a fixed delay before each volume to maintain steady-state conditions, independent of previous spin history. This concept allows for variable inter-volume delays and thus for flexible stimulus design in auditory fMRI. The purpose was to compare the signal stability of SEPIMS and conventional sparse EPI (CS-EPI). METHODS The saturation module comprises two non-selective adiabatic saturation pulses. The efficiency of the saturation and its effect on the SEPIMS signal stability is tested in vitro and in vivo. RESULTS Data show that SEPIMS yields the same signal stability as CS-EPI, even for extreme variations between inter-volume delay durations. However, dual saturation pulses are required to achieve sufficiently high saturation efficiency in compartments with long T1 values. Importantly, spoiler gradient pulses after the EPI readout have to be optimized to avoid eddy-current-induced image distortions. CONCLUSION The proposed SEPIMS sequence maintains high signal stability in the presence of variable inter-volume durations, thus allowing for flexible stimulus design.