QUBIC IV: Performance of TES Bolometers and Readout Electronics

2021 
A prototype version of the Q & U Bolometric Interferometer for Cosmology (QUBIC) underwent a campaign of testing in the laboratory at Astroparticle Physics and Cosmology in Paris. The detection chain is currently made of 256 NbSi Transition Edge Sensors cooled to 320 mK.The readout system is a 128:1 Time Domain Multiplexing scheme based on 128 SQUIDs cooled at 1K that are controlled and amplified by an SiGe Application Specific Integrated Circuit at 40 K. We report the performance of this readout chain and the characterization of the TESs. The readout system has been functionally tested and characterized in the lab and in QUBIC. The Low Noise Amplifier demonstrated a white noise level of 0.3 nV/sqrt(Hz). Characterizations of the QUBIC detectors and readout electronics includes the measurement of I-V curves, time constant and the Noise Equivalent Power. The QUBIC TES bolometer array has approximately 80% detectors within operational parameters. While still limited by microphonics from the pulse tubes and noise aliasing from readout system, the Noise Equivalent Power is about 2E-16 W/sqrt(Hz), enough for the demonstration of bolometric interferometry.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    6
    Citations
    NaN
    KQI
    []