Multipass-assisted dual-comb gas sensor for multi-species detection using a free-running fiber laser

2020 
We demonstrated a multipass-assisted dual-comb gas sensor for the simultaneous and sensitive detection of C2H2 and NH3 near 1.5 μm using a single free-running fiber laser. Two pulse trains generated from the shared cavity in the passively mode-locked fiber laser ensure the mutual coherence and common-mode noise rejection. One of the comb pulse trains passed through an open-path multipass cell with an effective path length of 4 m. A synchronized acquisition technique was adopted for more efficient interferogram acquisition compared with the conventional continuous acquisition method. At an acquisition time of 1.5 s, we obtained a signal-to-noise ratio of more than 600 by 1500 averages of the acquired spectrum of 0.1% C2H2. Pressure broadening of C2H2 and NH3 in the frequency range of 6503–6530 cm−1 between 0.1 and 1 bar were captured by the dual-comb sensor. Finally, the free-running fiber laser-based dual-comb sensor successfully analyzed the C2H2/NH3 mixtures, enabling real-time gas sensing without using complex optical and electronic feedback systems.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    33
    References
    6
    Citations
    NaN
    KQI
    []