Thermal calorimeters for high resolution X-ray spectroscopy

1993 
Abstract Thermal detection of individual X-ray photons by small (0.5×0.5 mm) calorimeters has been used to achieve an energy resolution as good as 7.5 eV FWHM for 6 keV X-rays. Such detectors should have interesting applications in X-ray astronomy as well as laboratory spectroscopy, and they promise a high tolerance for embedded sources. Ideally, it should be possible to improve the resolution greatly by making smaller detectors or operating them at lower temperatures than the 50–100 mK currently used. However, there appear to be fairly fundamental limitations when semiconductor thermistors are used as the thermometer. When trying to achieve energy resolution of 0.1% or better, fluctuations in the thermalization efficiency of the detector must also be considered, and this places additional restrictions on suitable detector materials.