Improved Detection of Microbubble Signals Using Power M-Mode Doppler

2003 
Background— Power motion-mode transcranial Doppler (TCD) (PMD) is a new, multigated technique that may simplify and enhance detection of embolus. We developed criteria for emboli detection using PMD. Then, we performed a blinded comparison of transcranial PMD with single-gate spectral TCD in TCD bubble study patients. Methods— Patients with right-to-left shunt as detected with standard TCD were selected for this study. The international emboli criteria for spectral TCD were used. We defined novel PMD criteria for detecting emboli signature on PMD as follows: (1) signature at least 3 dB higher than the highest spontaneous PMD display of background blood flow; (2) embolic signature reflects motion in one direction at a minimum spatial extent of 7.5 mm and temporal extent of 30 ms; (3) embolus must traverse a prespecified depth. Each study was blindly assessed for microbubble signals (MBS) count on either modality. Results— Thirty-six patients were included in the study. Mean age was 44.4 (SD 14.4), 50% were...