SKIN DOSE MAPPING IN INTERVENTIONAL CARDIOLOGY: A PRACTICAL SOLUTION

2020 
Numerous cases of radiation-induced tissue reactions following interventional cardiology (IC) procedures have been reported, resulting in the need for an optimized and personalized dosimetry. At present, there are many fluoroscopy units without Digital Imaging and Communications in Medicine (DICOM) Radiation Dose Structured Report globally installed. Many of these have not been updated yet, and may never be, therefore, the main objectives of this paper are to develop an offline skin dose mapping application, which uses DICOM headers for the peak skin dose (PSD) assessment and to compare the PSD assessment results to XR-RV3 Gafchromic film for common IC procedures. The mean deviation between the measured and the calculated PSD was 8.7 ± 26.3%. Simulated skin dose map showed good matching with XR-RV3 Gafchromic film. The skin dose mapping application presented in this paper is an elegant solution and a suitable alternative to XR-RV3 Gafchromic film.