Effect of Bipolar Electrode Spacing on Phrenic Nerve Stimulation and Left Ventricular Pacing ThresholdsClinical Perspective

2012 
Background— Phrenic nerve stimulation (PNS) is a common complication of cardiac resynchronization therapy when left ventricular (LV) pacing occurs via a coronary vein. The purpose of this study was to evaluate the effects of bipolar electrode spacing on PNS and LV pacing thresholds. Methods and Results— Electrophysiology catheters with standard (2 mm-5 mm-2 mm) or modified (1 mm-5 mm-1 mm) interelectrode spacing was, respectively, inserted in a posterior/lateral cardiac vein in a randomized order in 6 anesthetized dogs via jugular access. The phrenic nerve was dissected via a left minithoracotomy and repositioned over the vein as close as possible to one of the electrodes. The presence of PNS was verified (ie, PNS threshold 0.05). Compared with a standard bipolar electrode spacing of 20 mm for LV leads, 1 and 2 mm bipolar electrode spacing resulted in a PNS threshold increase of 5.5±2.2 V ( P =0.003) and 2.8±1.7 V ( P <0.001), respectively. Similarly, PNS threshold increased by 6.5±3.7 V with 1 mm and by 3.8±1.9 V with 2 mm bipolar pacing (both P <0.001), compared with unipolar pacing. Conclusions— This study suggests that reducing LV bipolar electrode spacing from the standard 20 mm to 1 or 2 mm may significantly increase the PNS threshold without compromising LV pacing thresholds.