This work is devoted to the differential die-away technique that is widely used for active detection of fissile materials via pulsed neutron generators. The technique allows direct detection of milligram quantities of uranium-235 and plutonium-239 in objects with volumes up to several cubic meters. Our group has demonstrated this technique, creating a special installation based on the commercially produced ING-07T pulsed neutron generator. The installation includes eight proportional 3He-counters mounted inside a polyethylene moderator with a cadmium filter, as well as a polyethylene chamber into which a 70-liter container is loaded for inspection. Preliminary testing showed that the minimum detectable mass of unshielded uranium-235 is ∼3 mg, using a 5.108 n/s neutron yield and 8 min measurement time. When the container is filled with neutron absorbing materials, e.g., iron, the minimum detectable mass increases to ∼30 mg. Use of borated screens further increases the minimum mass that can be detected. The tested installation and/or its modifications can be used for control and detection of fissile materials in various applications from luggage inspection to control containers with nuclear fuel cycle radioactive wastes.