Effect of increased vacuum and aspiration rates on phacoemulsification efficiency

2015 
Purpose To evaluate the effect of vacuum and aspiration rates on phacoemulsification efficiency. Setting John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, USA. Design Experimental study. Methods Formalin-soaked porcine lenses were divided into 2.0 mm cubes, and 0.9 mm 30-degree beveled 20-degree bent tips were used with micropulse ultrasound (US) (6 milliseconds on and 6 milliseconds off) and a peristaltic flow system. Vacuum levels were tested at 200, 300, 400, and 500 mm Hg, and aspiration rates were tested at 20, 35, and 50 mL/min. Efficiency (time to lens removal) and chatter (number of lens fragment repulsions from the tip) were determined. Results Increasing vacuum increased efficiency only when going from 200 mm Hg to higher vacuum levels. Increasing aspiration increased efficiency at all points measured (25 mL/min versus 35 mL/min, P P = .012; 25 mL/min versus 50 mL/min, P Conclusions Vacuum improved efficiency only up to 300 mm Hg and was more dependent on increasing flow. Similarly, chatter correlated with 200 mm Hg vacuum only and was more correlated with flow. Limitations of this study include use of only 1 US power modulation and hard nuclear material. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.