Magnetic field optimization and design of a superconducting neutron Wollaston prism

2016 
We present finite element simulations of a superconducting magnetic Wollaston prism (WP) for neutron scattering with high encoding efficiency and low Larmor phase aberrations. To achieve this, we develop and quantify the design criteria. The validation of simulation tools used for this work are investigated by using two software packages: RADIA and MagNet©. Based on the optimization criteria, various possible configurations of WP are explored with MagNet, from which the best configuration is chosen for further optimization. To optimize the best configuration, the influence of various physical parameters is investigated, including the dimensions, shapes and arrangements of components of the device. The optimum WP was built and measured at both pulsed and constant wavelength neutron sources. In flipping mode, a neutron spin flipping efficiency of ~98.5% was measured independent of neutron wavelength and applied current. In a precession mode, measurements showed a highly linear Larmor phase variation along the horizontal direction with low depolarization. Simulations of the device agree well with the experimental measurements. Possible applications of the device are also discussed.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    8
    Citations
    NaN
    KQI
    []