Magnetic resonance jet velocity mapping in mitral and aortic valve stenosis

1993 
BACKGROUNDMagnetic resonance (MR) phase-shift velocity mapping is an established method for measurement of nonturbulent intravascular flow. Shortening the echo time of the MR sequence to 3.6 msec allowed application of the technique to turbulent jet flow. The objective of this study was validation of MR jet velocity mapping in patients with cardiac valve stenosis.METHODS AND RESULTSWe used a 0.5-T Picker MR machine to measure peak poststenotic jet velocity in 15 consecutive patients recruited with known valve disease (six mitral stenosis, three of these restudied after valvoplasty, and 11 aortic stenosis). On the same day as the MR study, these patients underwent independent Doppler echocardiographic measurement of peak jet velocity. The results of 10 further MR investigations of aortic stenosis are also reported and compared with Doppler studies performed within 6 months. Of the 29 MR studies, 28 (97%) produced interpretable velocity maps, the one failure being attributed to misplacement of the imaging s...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    32
    References
    165
    Citations
    NaN
    KQI
    []