Ultra-low-dose thoracic CT with model-based iterative reconstruction (MBIR) in cystic fibrosis patients undergoing treatment with cystic fibrosis transmembrane conductance regulators (CFTR).

2021 
AIM To assess the utility of a volumetric low-dose computed tomography (CT) thorax (LDCTT) protocol at a dose equivalent to a posteroanterior (PA) and lateral chest radiograph for surveillance of cystic fibrosis (CF) patients. MATERIALS AND METHODS A prospective study was undertaken of 19 adult patients with CF that proceeded to LDCTT at 12 and 24 months following initiation of ivacaftor. A previously validated seven-section, low-dose axial CT protocol was used for the 12-month study. A volumetric LDCTT protocol was developed for the 24-month study and reconstructed with hybrid iterative reconstruction (LD-ASIR) and pure iterative reconstruction (model-based IR [LD-MBIR]). Radiation dose was recorded for each scan. Image quality was assessed quantitatively and qualitatively, and disease severity was assessed using a modified Bhalla score. Statistical analysis was performed and p-values of RESULTS Volumetric LD-MBIR studies were acquired at a lower radiation dose than the seven-section studies (0.08 ± 0.01 versus 0.10 ± 0.02 mSv; p=0.02). LD-MBIR and seven-section ASIR images had significantly lower levels of image noise compared with LD-ASIR images (p CONCLUSIONS The use of a volumetric LDCTT protocol (reconstructed with pure IR) enabled acquisition of diagnostic quality CT images, which were considered extremely useful for surveillance of CF patients, at a dose equivalent to a PA and lateral chest radiograph.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    28
    References
    1
    Citations
    NaN
    KQI
    []