A new, noninvasive method to monitor the aging of bioprosthetic valves in the mitral position☆

1992 
Abstract Bioprosthetic valves undergo a tissue degeneration of unpredictable onset and amount. This process alters the structure and function of the valve and consequently shortens its lifespan. The echocardiographic technique usually used in the follow-up of these patients does not provide accurate information concerning the amount of prosthesis tissue degeneration. A new technique has been developed based on the spectral analysis of the first heart sound, which enables the evaluation of prosthetic leaflet stiffness. The Young's modulus (E) and stress (s) of the valve leaflets were derived as functions of the inner diameter of the heterograft and its primary vibration frequency, which can be obtained from the frequency spectrum of the first heart sound. Thirty-six patients with a mitral bioprosthetic valve were studied. Fifteen had thickening or calcification, or both, of the valvular leaflets at echocardiographic examination. In patients with a normal valve, E and s showed a good correlation with the duration of implantation (r = 0.909, p