Battery Malfunction of a Leadless Cardiac Pacemaker: Worrisome Single-Center Experience

2018 
Leadless cardiac pacemaker (LCP) therapy has been established clinically as a feasible and safe alternative to conventional transvenous pacemaker therapy for patients with an indication for single-chamber right-ventricular pacing. 1-3 However, reports on loss of telemetry and pacing output due to abrupt battery failure called the safety of one of the two commercially available systems seriously into question. The initial battery advisory with the Nanostim™ LCP was issued by the manufacturer in October 2016, who instantly called a global stop to Nanostim™ implants. To this day, similar battery issues have not been described for the Micra™ transcatheter pacing system. Therefore, we thought to analyze the long-term pacemaker performance and rate of battery malfunction of the Nanostim™ LCP in our patient population.