Evaluation of a Device Combining an Inferior Vena Cava Filter and a Central Venous Catheter for Preventing Pulmonary Embolism Among Critically Ill Trauma Patients

2017 
Abstract Purpose To evaluate efficacy and safety of a novel device that combines an inferior vena cava (IVC) filter and central venous catheter (CVC) for prevention of pulmonary embolism (PE) in critically ill patients. Materials and Methods In a multicenter, prospective, single-arm clinical trial, the device was inserted at the bedside without fluoroscopy and subsequently retrieved before transfer from the intensive care unit (ICU). The primary efficacy endpoint was freedom from clinically significant PE or fatal PE 72 hours after device removal or discharge, whichever occurred first. Secondary endpoints were incidence of acute proximal deep venous thrombosis (DVT), catheter-related thrombosis, catheter-related bloodstream infections, major bleeding events, and clinically significant thrombus (occupying > 25% of volume of filter) detected by cavography before retrieval. Results The device was placed in 163 critically ill patients with contraindications to anticoagulation; 151 (93%) were critically ill trauma patients, 129 (85%) had head or spine trauma, and 102 (79%) had intracranial bleeding. The primary efficacy endpoint was achieved for all 163 (100%) patients (95% confidence interval [CI], 97.8%–100%, P Conclusions This novel device prevented clinically significant and fatal PE among critically ill trauma patients with low risk of complications.